Banner

Measurement of Luminous Intensity Using Filtered Trap Detector MS Khalil, FM E-Sharkawy, MG EL-Shaarawy, N Shash, A Mahmoud International Journal of Optics and Applications 2 (4), 43-47

International Journal of Optics and Applications • 2012
Back
Publication Information
Authors MS Khalil, FM E-Sharkawy, MG EL-Shaarawy, N Shash, A Mahmoud
Keywords Not Available
Journal International Journal of Optics and Applications
Publisher Scientific & Academic Publishing
Volume 2
Issue 4
Pages 43-47
publication.type International
Paper Link Not Available
Supplementary Materials Not Available
Abstract
The work described in this paper is a part of the process of lamp and photometer calibration
to obtain a new luminous-intensity scale with a lower uncertainty. The procedure and the
results of realizing the photometric scale with a standardized detector at the National
Institute for Standards (NIS) are described. This method enables photometric values to be
derived from absolute spectroradiometric measurements by exact computation. A filtered
trap detector has been characterized and used to measure the absolute spectral power ...