publication name Measurement of Luminous Intensity Using Filtered Trap Detector MS Khalil, FM E-Sharkawy, MG EL-Shaarawy, N Shash, A Mahmoud International Journal of Optics and Applications 2 (4), 43-47
Authors MS Khalil, FM E-Sharkawy, MG EL-Shaarawy, N Shash, A Mahmoud
year 2012
keywords
journal International Journal of Optics and Applications
volume 2
issue 4
pages 43-47
publisher Scientific & Academic Publishing
Local/International International
Paper Link Not Available
Full paper download
Supplementary materials Not Available
Abstract

The work described in this paper is a part of the process of lamp and photometer calibration to obtain a new luminous-intensity scale with a lower uncertainty. The procedure and the results of realizing the photometric scale with a standardized detector at the National Institute for Standards (NIS) are described. This method enables photometric values to be derived from absolute spectroradiometric measurements by exact computation. A filtered trap detector has been characterized and used to measure the absolute spectral power ...