Characterization of a novel tellurite selective electrode.
• 1997
Publication Information
Authors
M.M. Zareh and A.S. Amin.
Keywords
Not Available
Journal
Not Available
Publisher
Not Available
Volume
Not Available
Issue
Not Available
Pages
Not Available
publication.type
Local
Paper Link
Not Available
Supplementary Materials
Not Available
Abstract
Tellurium was determined by introducing a solid membrane for the selective determination of tellurite ion. An electrode composed of a compressed disk containing a mixture of HgTeO3and Hg2Cl2was found to give a typical Nernstian response of 29 mV/concentration decade for the concentration range 10−5to 10−1M). The connection was made by Hg metal instead of aqueous inner filling. The optimum pH was 3.5–10.5. The selectivity of the electrode toward several inorganic and organic anions was estimated. The electrode was applied to the determination of tellurium in binary mixtures with a mean recovery of 98.99% and relative standard deviation range of 0.19–0.51%.
Staff Members - Benha University