Central Serous Retinopathy. Observations and a new explanation for the fluorescein angiographic findings.
• 1996
Publication Information
Authors
El-Deeb,M.A ; and Higazy,M.T
Keywords
Not Available
Journal
Not Available
Publisher
Not Available
Volume
Not Available
Issue
Not Available
Pages
Not Available
publication.type
International
Paper Link
Not Available
Supplementary Materials
Not Available
Abstract
Purpose: : To study the effect corneal biomechanical properties effect on 10P measurements, and compare this to Goldman applanation tonometer (GAT) the gold standard.
Methods: : An observational cross-sectional clinical study, 180 eyes of 94 normal participants were included. All participants underwent IOP measurement by ORA and GAT. Corneal compensated 10P (10Pcc), Goldman related 10Pg , corneal hysteresis (CH) and Corneal resistance factor (CRF) were recorded. Statistical analysis of the data was performed
Results: : Mean 10P measured by GAT was 14.03±3.00mmHg. ORA measurements for 10Pg was 14.55±2.95mmHg.10Pcc ORA measurements was 16.3±3.5mmHg. The recorded values of the GAT and ORA-I0Pg were nearly similar indicating accuracy of ORA in measuring lOP. Mean difference between 10Pcc and 10Pg was 4.7mmHg , and this showed a significant negative correlation with CH. (r= -0.98)
Conclusions: : The biomechanical properties of the cornea (CH, CRF) are critical parameters that influence 10P measurements.
Methods: : An observational cross-sectional clinical study, 180 eyes of 94 normal participants were included. All participants underwent IOP measurement by ORA and GAT. Corneal compensated 10P (10Pcc), Goldman related 10Pg , corneal hysteresis (CH) and Corneal resistance factor (CRF) were recorded. Statistical analysis of the data was performed
Results: : Mean 10P measured by GAT was 14.03±3.00mmHg. ORA measurements for 10Pg was 14.55±2.95mmHg.10Pcc ORA measurements was 16.3±3.5mmHg. The recorded values of the GAT and ORA-I0Pg were nearly similar indicating accuracy of ORA in measuring lOP. Mean difference between 10Pcc and 10Pg was 4.7mmHg , and this showed a significant negative correlation with CH. (r= -0.98)
Conclusions: : The biomechanical properties of the cornea (CH, CRF) are critical parameters that influence 10P measurements.